Event Notification Report for June 19, 2023
U.S. Nuclear Regulatory Commission
Operations Center
EVENT REPORTS FOR
06/18/2023 - 06/19/2023
Non-Power Reactor
Event Number: 56602
Rep Org: Univ Of New Mexico (NEWM)
Licensee: University Of New Mexico
Region: 0
City: Albuquerque State: NM
County: Bernalillo
License #: R-102
Agreement: Y
Docket: 05000252
NRC Notified By: Carl Willis
HQ OPS Officer: Eric Simpson
Licensee: University Of New Mexico
Region: 0
City: Albuquerque State: NM
County: Bernalillo
License #: R-102
Agreement: Y
Docket: 05000252
NRC Notified By: Carl Willis
HQ OPS Officer: Eric Simpson
Notification Date: 06/30/2023
Notification Time: 14:10 [ET]
Event Date: 06/19/2023
Event Time: 23:59 [MDT]
Last Update Date: 06/30/2023
Notification Time: 14:10 [ET]
Event Date: 06/19/2023
Event Time: 23:59 [MDT]
Last Update Date: 06/30/2023
Emergency Class: Non Emergency
10 CFR Section:
10 CFR Section:
Person (Organization):
Helvenston, Edward (NRR)
Waugh, Andrew (NRR)
Helvenston, Edward (NRR)
Waugh, Andrew (NRR)
TECHNICAL SPECIFICATION VIOLATION
The following information is a summary provided by the licensee via phone and email:
The University of New Mexico reported a violation of its non-power reactor technical specification (TS) 1.1.36(c) required alarm testing frequency. The alarm testing requirement requires that the site alarms be tested on a semi-annual basis, not to exceed 7.5 months. According to the licensee, the last successful test of the New Mexico non-power reactor alarm system occurred on October 27, 2022, which is in excess of the 7.5 month requirement. The reactor was at zero percent power when the TS violation was discovered and continues to be at zero percent power. Reactor operations are suspended until the issue is resolved.
The following information is a summary provided by the licensee via phone and email:
The University of New Mexico reported a violation of its non-power reactor technical specification (TS) 1.1.36(c) required alarm testing frequency. The alarm testing requirement requires that the site alarms be tested on a semi-annual basis, not to exceed 7.5 months. According to the licensee, the last successful test of the New Mexico non-power reactor alarm system occurred on October 27, 2022, which is in excess of the 7.5 month requirement. The reactor was at zero percent power when the TS violation was discovered and continues to be at zero percent power. Reactor operations are suspended until the issue is resolved.
Agreement State
Event Number: 56581
Rep Org: California Radiation Control Prgm
Licensee: Pavement Engineering, Inc.
Region: 4
City: Sacramento State: CA
County:
License #: 8183-07
Agreement: Y
Docket:
NRC Notified By: Robert Greger
HQ OPS Officer: Brian P. Smith
Licensee: Pavement Engineering, Inc.
Region: 4
City: Sacramento State: CA
County:
License #: 8183-07
Agreement: Y
Docket:
NRC Notified By: Robert Greger
HQ OPS Officer: Brian P. Smith
Notification Date: 06/19/2023
Notification Time: 18:32 [ET]
Event Date: 06/19/2023
Event Time: 01:00 [PDT]
Last Update Date: 06/19/2023
Notification Time: 18:32 [ET]
Event Date: 06/19/2023
Event Time: 01:00 [PDT]
Last Update Date: 06/19/2023
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
10 CFR Section:
Agreement State
Person (Organization):
Warnick, Greg (R4DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)
ILTAB, (EMAIL) (EMAIL)
CNSNS (Mexico), - (FAX)
Warnick, Greg (R4DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)
ILTAB, (EMAIL) (EMAIL)
CNSNS (Mexico), - (FAX)
AGREEMENT STATE REPORT - STOLEN MOISTURE DENSITY GAUGE
The following report was received by the California Department of Public Health (the Department) via email:
"On June 19, 2023, the licensee notified the Department that at 0100 [PDT] one nuclear gauge (Troxler 3430, SN 19706) containing 8 mCi Cs-137 and 40 mCi Am-241/Be was stolen from a locked vehicle that was parked in an apartment complex. The gauge was stored in its case that was chained and locked to the pick up truck bed. The gauge was not visible as the bed of the truck was fitted with a locking cover. The cover had been broken open and the locks securing the gauge were cut. Sacramento Police were notified and an electronic report was filed.
"The Department will investigate."
California Event Number: 061923
THIS MATERIAL EVENT CONTAINS A 'Less than Cat 3' LEVEL OF RADIOACTIVE MATERIAL
Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks. For additional information go to http://www-pub.iaea.org/MTCD/publications/PDF/Pub1227_web.pdf
The following report was received by the California Department of Public Health (the Department) via email:
"On June 19, 2023, the licensee notified the Department that at 0100 [PDT] one nuclear gauge (Troxler 3430, SN 19706) containing 8 mCi Cs-137 and 40 mCi Am-241/Be was stolen from a locked vehicle that was parked in an apartment complex. The gauge was stored in its case that was chained and locked to the pick up truck bed. The gauge was not visible as the bed of the truck was fitted with a locking cover. The cover had been broken open and the locks securing the gauge were cut. Sacramento Police were notified and an electronic report was filed.
"The Department will investigate."
California Event Number: 061923
THIS MATERIAL EVENT CONTAINS A 'Less than Cat 3' LEVEL OF RADIOACTIVE MATERIAL
Sources that are "Less than IAEA Category 3 sources," are either sources that are very unlikely to cause permanent injury to individuals or contain a very small amount of radioactive material that would not cause any permanent injury. Some of these sources, such as moisture density gauges or thickness gauges that are Category 4, the amount of unshielded radioactive material, if not safely managed or securely protected, could possibly - although it is unlikely - temporarily injure someone who handled it or were otherwise in contact with it, or who were close to it for a period of many weeks. For additional information go to http://www-pub.iaea.org/MTCD/publications/PDF/Pub1227_web.pdf
Agreement State
Event Number: 56582
Rep Org: Arkansas Department of Health
Licensee: Jefferson Regional Medical Center
Region: 4
City: Pine Bluff State: AR
County:
License #: ARK-0623-02120
Agreement: Y
Docket:
NRC Notified By: Susan Elliott
HQ OPS Officer: Ernest West
Licensee: Jefferson Regional Medical Center
Region: 4
City: Pine Bluff State: AR
County:
License #: ARK-0623-02120
Agreement: Y
Docket:
NRC Notified By: Susan Elliott
HQ OPS Officer: Ernest West
Notification Date: 06/20/2023
Notification Time: 10:41 [ET]
Event Date: 06/19/2023
Event Time: 10:30 [CDT]
Last Update Date: 06/20/2023
Notification Time: 10:41 [ET]
Event Date: 06/19/2023
Event Time: 10:30 [CDT]
Last Update Date: 06/20/2023
Emergency Class: Non Emergency
10 CFR Section:
Agreement State
10 CFR Section:
Agreement State
Person (Organization):
Kellar, Ray (R4DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)
Kellar, Ray (R4DO)
NMSS_EVENTS_NOTIFICATION (EMAIL)
AGREEMENT STATE REPORT - PATIENT UNDERDOSE
The following information was provided by the Arkansas Department of Health via email:
"On June 19, 2023, at approximately 1300 CDT, the licensee contacted the Arkansas Department of Health to report a medical event from Jefferson Regional Medical Center, Pine Bluff, Arkansas. The Radiation Safety Officer, Imaging Services Compliance Specialist, called to report a mishap with their first procedures with Y-90. It appears that their patient was to have received 19.8 Gray. However, due to equipment or patient issues the patient received only 6% (i.e. 2.6 Gray). The original injection was to have been 68.5 mCi.
"The report from the licensee is as follows: This is a termination of the Y-90 therapy due to stasis of the administration. 'Based on dosimeter readings of the Y-90 vial, and Geiger-Mueller meter measurements of the equipment, the microspheres were blocked in a tubing connector and did not fully reach the patient and/or organ. There was no spillage or contamination of microspheres from the Y-90 dose. The tubing and catheter were placed in appropriate shielding for long term decay in storage. Below and attached are the calculations we received from the lead Nuclear Medicine technologist and Therasphere representative.'
"Written Directive: 68.6 mCi
"Administration: 64.9 mCi at 1030 [CDT]
"Start: 1138 [CDT]
"Stasis Termination: 1209 [CDT]
"Dose to patient: 3.6 mCi"
AR event number: ARK-2023-004
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.
The following information was provided by the Arkansas Department of Health via email:
"On June 19, 2023, at approximately 1300 CDT, the licensee contacted the Arkansas Department of Health to report a medical event from Jefferson Regional Medical Center, Pine Bluff, Arkansas. The Radiation Safety Officer, Imaging Services Compliance Specialist, called to report a mishap with their first procedures with Y-90. It appears that their patient was to have received 19.8 Gray. However, due to equipment or patient issues the patient received only 6% (i.e. 2.6 Gray). The original injection was to have been 68.5 mCi.
"The report from the licensee is as follows: This is a termination of the Y-90 therapy due to stasis of the administration. 'Based on dosimeter readings of the Y-90 vial, and Geiger-Mueller meter measurements of the equipment, the microspheres were blocked in a tubing connector and did not fully reach the patient and/or organ. There was no spillage or contamination of microspheres from the Y-90 dose. The tubing and catheter were placed in appropriate shielding for long term decay in storage. Below and attached are the calculations we received from the lead Nuclear Medicine technologist and Therasphere representative.'
"Written Directive: 68.6 mCi
"Administration: 64.9 mCi at 1030 [CDT]
"Start: 1138 [CDT]
"Stasis Termination: 1209 [CDT]
"Dose to patient: 3.6 mCi"
AR event number: ARK-2023-004
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.